Is it safe to remove and insert the SD memory card during measurement?

The SD memory card can be replaced during measurement by pressing the [FILE] key and selecting “Remove/Exchange SD Card”.
However, the replacement should be done within 10 minutes.
Although the files will be separated when the SD memory card is replaced, there will be no data loss between files as long as the replacement is done within 10 minutes.
When replacing an SD memory card using the “Remove/Exchange SD Card” function during data recording, make sure that the “Device Access Display” is green, and then press the [ENTER] key.
When the “Device Access Display” is red, the SD memory card is in the process of being accessed. If the [ENTER] key is pressed during access, subsequent data recording may not be performed properly.